Firmware Version 04.02.25 and above contain a bug that will cause the transmitter to lockup and not function if the displayed value for flow rate is greater than five digits (expressed as an exponential number; for example, 2.4e05). This is most likely to be seen with large diameter pipes and units of measure that result in large numbers.

<h1>Bug in Firmware Versions 04.02.25 and Above Causes Transmitter Lockup</h1>

Firmware Version 04.02.25 and above contain a bug that will cause the transmitter to lockup and not function if the displayed value for flow rate is greater than five digits (expressed as an exponential number; for example, 2.4e05). This is most likely to be seen with large diameter pipes and units of measure that result in large numbers.

Categories: Technical Bulletin
Tags: English
Product: SONARtrac
filename: TB021 Rev. 1
et_enqueued_post_fonts: a:3:{s:6:"family";a:3:{s:12:"et-gf-oswald";s:34:"Oswald:200,300,regular,500,600,700";s:13:"et-gf-poppins";s:134:"Poppins:100,100italic,200,200italic,300,300italic,regular,italic,500,500italic,600,600italic,700,700italic,800,800italic,900,900italic";s:11:"et-gf-actor";s:13:"Actor:regular";}s:6:"subset";a:2:{i:0;s:5:"latin";i:1;s:9:"latin-ext";}s:9:"cache_key";s:74:"{"gph":0,"divi":"4.27.4","wp":"6.8.2","enable_all_character_sets":"false"}";}