by Aimee Hernandez | Feb 15, 2024
<h1>Transmitter Problem When Using Totalizer with Custom Units</h1> A problem with the SONARtrac transmitter ‘totalizer’ function has been discovered. When a ‘USER’ unit (customized volumetric unit based on an existing standard transmitter defined unit) is...
by Aimee Hernandez | Feb 15, 2024
<h1>Transmitter Problem When Using Fixed Blocks</h1> If a ‘SENSOR TEST’ is performed on a flow transmitter configured to use ‘FIXED BLOCKS’ (seconds of data) or ‘FIXED ALL’ (seconds of data and fixed frequency search limits), the transmitter will use only...
by Aimee Hernandez | Feb 15, 2024
<h1>Transmitter Rebooting When Using Field Utility Programs</h1> Setting the temperature input (Sensor #1 or #2) to other than fixed when the pressure input is set to fixed causes the transmitter to reboot when using field utility programs (Flow Dev,...
by Aimee Hernandez | Feb 15, 2024
<h1>Spike Filter Problems</h1> Under certain circumstances, issues have been identified with released versions of the spike filter that can cause improper functionality.
by Aimee Hernandez | Feb 15, 2024
<h1>Transmitter Totalizer Problem</h1> A problem with the transmitter totalizer function has been found. The problem exists in all Firmware Releases up to and including 03.03.26.