by Aimee Hernandez | Feb 15, 2024
<h1>Preamplifier Communication Errors</h1> There have been reports of preamplifier communication errors on some systems. The problem shows up as a failed sensor test, or inability to adjust the gain, or sensor temperature not updating....
by Aimee Hernandez | Feb 15, 2024
<h1>Alternate Transmitter Display Default Contrast Setting</h1> The unavailability of the display normally used in our transmitter has required us to use an alternate one. About 50 transmitters will be equipped with this alternate display. This is expected...
by Aimee Hernandez | Feb 15, 2024
<h1>Incomplete Snapshot Data Sets</h1> Sometimes when taking a Snapshot data set the Raw Data portion of the data set will not be saved to the file. The problem may be exclusive to Snapshots collected using the USB port and memory stick. This problem is...
by Aimee Hernandez | Feb 15, 2024
<h1>Gauge Block and Screw Size Table</h1> The referenced table lists the sensor band part number, spacer gauge tool, sensor band screw hex wrench size and band attachment rail size. This table has been revised to include gauges and sensors not currently...
by Aimee Hernandez | Feb 15, 2024
<h1>Bug in Firmware Versions 04.02.25 and Above Causes Transmitter Lockup</h1> Firmware Version 04.02.25 and above contain a bug that will cause the transmitter to lockup and not function if the displayed value for flow rate is greater than five digits...